Compare DHT22, DHT11 and Sensirion SHT71

Page Contents

 

Introduction

My previously published results compared six AM2302 (a.k.a. DHT22, RHT03 and I use the names interchangably throughout) hygrometers. Here I repeat that experiment using the same apparatus and techniques but replace two of the sensors with alternative models, a DHT11 and a Sensirion SHT71. The objectives of the earlier work were to establish whether a sensor as cheap as the AM2302/DHT22 could live up to their claimed accuracy. My conclusion was that in my experiments they did not, but they did deliver surprisingly good performance and very good value for money for most non-safety-critical, domestic DIY projects. The next obvious question is whether my tests would show a more expensive device to be better. Since I was re-running the experiment I also included the even cheaper and lower specification DHT11.

Data sheets for the DHT11 and DHT22 devices tend to be brief. The numbers in the following table appear on datasheets and are typically quoted by retailers. The Sensirion datasheet on the other hand is detailed and comprehensive providing accuracy as a function of humidity as well as details of recommended calibration and linearization procedures. Note how Sensirion's absolute accuracy claims are less strict and more believable than those normally quoted for the DHT devices.

Manufacturers' Specification

 

AM2302 / DHT22 DHT11 SHT71
Range 0-100% 20-90% 0-100%
Absolute accuracy ±2% ±5% ±3% (20<RH<80)
±5% (RH<20, RH>80)
Repeatability ±1% ±1% ±0.1%
Long term stability ±0.5% per year ±1% per year <0.5% per year
Typical street price US$ 4-10 US$ 1-5 US$ 30-50

UPDATE: Since first writing this page a much improved data sheet has recently appeared which contains both clearer translations to English and more detailed specifications and plots. It still claims a typical accuracy of ±2%, but does now show the accuracy degrading to ±5% at the two extreme limts, <10% and >90%

Accurately and repeatably measuring relative humidity is notoriously tricky. The procedures used here were developed over a period of about a year and are detailed on my DHT22/AM2302 calibration page. I am no expert in hygrometers. I just devised the best experiment I could based on my reading of several papers on the topic and using a few items of household equipment I had lying around.

 

The Devices and Test Apparatus

The AM2302/DHT22 devices are the same units as I used previously. They are A,B,D,E and F from my previous write-up. Though five are mentioned, only four were under test at any one time. Sensor B failed during the experiment and was replaced by E. I have added a DHT11 and a Sensirion SHT71.

Apparatus setup is as previously described. All sensors were powered from a 5V d.c. switching power supply. New software needed to be added to microcontroller to read the Sensirion device and was based on Markus Schatzl and Carl Jackson's Sensirion Arduino library.

The reference calibration sources are still the same too, eleven saturated salt solutions and distilled water. Data were all collected in a similar manner to before with the sensors being allowed to stabilise for a few hours with each solution.

Quality of build

At more than ten times the price the SHT71 is unsuprisingly far superior to the others. It is both smaller and feels more solid. The gold plated Cu/Be alloy pins are very robust in comparison to the DHT22 on which the pins honestly feel like they are made of thick aluminium foil. Note that the SHT71 has 1.27mm separation pins which does make it less easy to hook up to common hobbyist 2.54mm Arduinos and bread-boards. I mounted mine in a 2.54mm header block for easy handling.

Image of SHT71 alongsid DHT22
Figure 1. Comparison of the packaging of the SHT71 compared to AM2302. DHT11 looks similar to the AM2302. Though somewhat smaller, it also has 0.1" separation pins. The SHT71 has 0.05" separation pins.

Response Speed

The SHT71 consistently responded to changes the fastest, registering a change in a few seconds. The DHT22/AM2302 appears to take about 30sec and the DHT11 can take a couple of minutes. However, the DHT22 caches a reading in memory and returns it whenever a value is next requested. Since I am only sampling every 30sec, the DHT22 values are always from 30sec ago which is why the step change in Figure 2 lags 30sec behind the SHT71.

All sensors (including the SHT71) can take several hours to fully stabilize at high humidities. Though some of this may be the device, I suspect it genuinely takes several hours to equilibrate and saturate the air inside the jar after a swap. Still, the relative fact that the SHT71 is the fastest and DHT11 the slowest is obviously real since they are all together measuring the same air.

A plot of humidity values as a function of time.
Figure 2. A typical example of humidity measurements as the sample solution is changed. Output values from the SHT71 consistently respond on the first reading after the change and shows an abrupt step change. The DHT22s typical responds one sample later because it returns cached, not new, measurements. The DHT11 shows some sort of reponse with a minute or two, but can take a while to gradually drift towards the new measurement.

Results

Part 1: As a Function of Humidity

First we look at the varying response of the sensor to different reference humidities, all measured at a single fixed temperature.

Compound Ref. Measured RH %
RH % A B D F SHT71 DHT11
NaOH 6.8 9.7 12.5 10.2 8.4 12.7 31.8
LiCl 11.2 14.0 15.8 14.8 12.9 16.6 31.9
MgCl 32.8 31.6 29.2 33.9 31.4 35.4 38.9
K2CO3 42.6 41.4 37.0 45.3 42.6 45.4 46.5
NaBr 56.6 54.4 46.5 59.0 56.7 57.4 57.9
NH4NO3 59.4 57.1 48.9 61.9 59.7 60.7 61.9
KI 67.9 65.0 54.6 71.8 69.1 68.4 70.3
NaCl 75.3 71.8 60.1 80.3 78.9 75.8 80.3
NH4SO4 79.9 75.9 63.4 85.7 84.6 80.1 86.3
KCl 84.0 79.1 65.6 89.6 91.3 83.8 89.6
K2NO3 91.7 87.4 71.1 98.0 - 91.6 91.0
H2O 100.0 96.4 77.8 - - 98.1 92.0
Experimental results at run 2 (August 2014), taken at 30°C. 'Reference RH' is the expected value taken from published literature interpolated to the temperature at the time the measurement was obtained. 'Measured RH' are the values as returned from the DHT22, DHT11 and SHT71 devices. In the case of SHT71, these numbers have already been temperature corrected and linearized using the default paramaters from the manufacturer's datasheet. Each value is average of two or three measurements obtained over a period of two weeks. For sensors A,B,D and F these are new measurements and may be compared to previous data obtained from the same sensors. Temperature is entirely defined by the readings from the devices themselves without any external calibration.

 

Compound Ref. Measured RH %
RH % A B D E F DHT11 SHT71
NaOH 7.3 9.7 9.7 8.4 9.4 7.8 35.8 12.6
LiCl 11.8 14.0 13.3 12.8 13.8 12.1 35.9 16.3
MgCl 33.1 33.3 31.0 31.9 32.7 30.9 38.9 35.5
K2CO3 43.4 44.1 41.7 42.8 45.4 41.6 48.6 45.3
NaBr 58.1 59.2 56.1 59.2 61.0 58.3 63.3 59.7
NH4NO3 64.7 64.1 61.0 63.9 65.4 64.1 67.4 64.2
KI 69.5 70.2 66.7 72.3 71.3 71.2 74.0 70.4
NaCl 75.3 76.4 72.2 79.0 76.4 79.3 82.4 76.2
NH4SO4 80.2 82.0 77.3 84.7 81.0 86.6 91.4 81.4
KCl 85.3 86.3 82.0 88.0 85.1 93.0 93.7 85.2
K2NO3 93.5 96.3 - 98.0 95.3 - 95.0 93.5
H2O 100.0 - - - - - - 98.7
Experimental results at run 3 (November 2014), taken at 22°C.

 

Plots of measured vs. reference humidity for the eight hygrometers
Figure 3. Plots of values read from hygrometer against the known reference humidity. Three epochs are shown. Green is run 1 from May 2014. Blues are run 2 from August 2014 and split into cyan fot those data known to be corrupted by B self-heating and dark blue for data thought to be good. Red are run 3 from November 2014.

 

Plots of (measured-reference) vs. reference humidity.
Figure 4. The same data as Figure 3 with the same quadratic polynomial fits are replotted but the vertical axis is now the difference between measured and reference values. These plots present the error that would result from using the factory calibrated values. For DHT11 and DHT22 these are read directly from the sensor without any re-calibration. For SHT71 these values have been temperature corrected and linearized using the default paramaters from the manufacturer's datasheet. Three epochs are shown. Green is run 1 from May 2014. Blues are run 2 from August 2014 and split into cyan for those data known to be corrupted by B self-heating and dark blue for data thought to be good. Red are run 3 from November 2014. Grey shaded area is the specification from datasheets.

 

Sensirion SHT71

This is the best of the sensors. It is the most linear, most stable over time and arguably the one with the smallest absolute deviations though cherry picking the best of the DHT22s, they are comparable. It may justify its cost if you have a need for that extra accuracy, and particularly reliabilty. For most everyday purposes the other sensors are probably adequate except for the gross inconsistency caused by sensor B's self-heating. Repeatability and consistency is where the SHT71 seems to win easily. Finer manufacturing tolerances and quality control are presumably what you are paying for with the more expensive devices. RMS scatter around the fit line is 2%RH, but this is only an estimate of the overall accuracy if the correction curve is applied and for as long as that correction curve remains unchanged. Note that that 2%RH scatter includes systematic errors in my apparatus as well as measurement errors in the sensors. The true humidity generated by each solution is only known to about 2%RH. For example all the sensors give 1–2% lower than expected readings for ammonium nitrate at 22°C, suggesting it is the reference data I am using that is in error rather then the sensors. Without my own correction curve, errors from the sensor after applying the manufacturer's default calibration from the datasheet are up to 5%. All my data points nearly remain within the shaded area of the manufacturer's specification.

DHT11

As specified on the data sheet, this device is of no use below 20% or above 90%, but then in terms of physical comfort, anything above 90% humidity feels the same, i.e., wet. Similarly at anything below 20% my lips start cracking so for many uses the difference between 5% and 15% may not be important. The repeatability (scatter of the data points) is markedly worse than all the other sensors (±5%) but within its valid range (20 < %RH < 90) its absolute calibration is almost as good as the DHT22s. A calibration curve is not justified by these data though a constant offset of about 4% would appear to improve reading accuracy. If the self-heating of sensor B was affecting the adjacent DH11 then the required offset could be slightly greater. A data run without the self-heating B was started, but then abandoned when I decided to no longer pursue use of this device.

DHT22 / AM2302

Sensor A Ignoring run 2 which was corrupted by faulty sensor B, this device looked good until just before the end of the experiment when it became the second of the six DHT22s to fail. When working, it consistently read 2% high.
Sensor B is highly problematic. During the second data run the device was faulty and running hot. The heat was also influencing its own local environment so it has little use as a measure of the surrounding ambient conditions. Even when not self-heating in run 3 its behaviour seems to have changed to some degree. This device has been scrapped.
Sensor C Only tested once during which its results were remarkably similar to the SHT71.
Sensor D has changed more than the specification allows, but is still tolerable with a 5% error or so. Its changes are not explained by local heating from sensor B. Applying any of the correction curves would improve the other measurements so it is showing some consistency, but it has clearly changed.
Sensor E looks good. Divergence at 100% could just be a couple of data logging errors in run 1 and if you were to ignore them it has remained very consistent.
Sensor F has changed little between measurements. Unfortunately it has the most aggressively curved of all the calibration curves, but it has at least remained reasonably constant. If I were applying a correction curve derived from the old data it would still be valid now.

Part 2: As a Function of Temperature

The above measurements were taken at fixed temperatures (30°C and 22°C). Next we look at how the sensors react over the range 10–40 °C. There are two effect to be disentangled. We wish to measure whether the sensors' response changes with temperature, but we know the humidity generated by the solutions is itself temperature sensitive. The 'reference values' are therefore no longer fixed constants, but temperature dependent slopes. DHT22 sensors A,D,E,F, the DHT11 and the SHT71 were tested with all the saturated solutions and plots for three are presented in Figures 5, 6 and 7. The compounds selected for inclusion here are:

Plots showing thermal dependence of sensor output. Measured humidity vs. temperature.
Figure 5. DHT22 sensors A,D,E,F, the DHT11 and the SHT71 tested with saturated sodium chloride over the temperature range 12–35°C. Over-plotted for comparison are values taken from the published literature. Though it reads 1% high, the SHT71 is only one to show the correct, temperature independent behaviour. All the DHT22s show humidity readings that rise with temperature, consistent with my earlier DHT22-only study

 

Plots showing thermal dependence of sensor output. Measured humidity vs. temperature.
Figure 6. DHT22 sensors A,D,E,F, the DHT11 and the SHT71 with saturated ammonium nitrate over the temperature range 10–35°C. Over-plotted for comparison are values taken from the published literature. Two data runs were obtained a couple of months apart. All the sensors correctly illustrate the fact that the ammonium nitrate solubility is strongly temperature dependent causing the humidity above the saturated solution to change with temperature. The SHT71 again best matches the reference data. Most sensors exhibit a 1-2% drift between the two data runs. Sensor E has remained very consistent over time though as with all the DHT devices, it again exhibits the tendency to read higher humidity than it should as the temperature increase. Sensor A failed between the two data runs and became the second of my six DHT22s to die. As with sensor B it exhibits the insidious behaviour of still producing sensible looking results which correlate correctly with temperature. It is simply offset by 15%.

Plots showing thermal dependence of sensor output. Measured humidity vs. temperature.
Figure 7. DHT22 sensors A,D,E,F, the DHT11 and the SHT71 with saturated magnesium chloride over the temperature range 10–32°C. Over-plotted for comparison are values taken from the published literature. Conclusions are consistent with previous measurements. The slope of the SHT71 again best matches the reference data meaning it has the smallest temperature related errors. The DHT22 outputs all have a tendency to increase with temperature. Though the temperature coefficient is wrong, the absolute calibration of sensor A is here the best of the lot. Sadly, shortly after taking these data it failed as illustrated in Figure 6.

These plots demonstrate again a point made repeatedly that these experiments are only as accurate as the availability of calibration references and the literature shows a considerable variation. Look for example at ammonium nitrate in Figure 6. My two data runs with the SHT71 show a systematic offset which is presumably sensor calibration drift, but that offset is only about the same same as the discrepancy between the published Wexler and O'Brien data sets.

The very obvious difference between Figures 5 and 6 demonstrates the success of the system setup. As with Figure 3 on the DHT22 report the primary conclusion of this is that we can clearly differentiate between sensor sensitivity changes and genuine environmental changes and our conclusions regarding sensor calibraton are valid, not experimental error.

Part 3: As a Function Simultaneously of Temperature and Humidity

Finally, if a sensor is to be used to measure humidity under a range of varying temperatures a full bivariate calibration is required. Such a calibration over the rather moderate temperature range 10 < °C < 35 is shown in Figure 8. The curves in Figure 4 are effectively cross sections through these surfaces.

3D surface plots representing the error as a surface in temperature-humidity space.
Figure 8. Surfaces showing deviation of the sensor output from the reference value as a function of both temperature and relative humidity. Values are for DHT11 and DHT22 are read straight from the devices. Values from SHT71 have been temperature corrected and linearized using the default paramaters from the manufacturer's datasheet. For DHT22 device A I include only data from before the sudden 15% shift so this plot represents how it performed for about 18months but is no longer valid.

Temperature Accuracy

At a correspondent's request I include a quick comparison of the temperature outputs. My apparatus contains no external reference against which to calibrate the temperature output and I therefore simply plot a direct comparison. Strictly this only shows they agree, not that they are all correct, but I don't think it is seriously in doubt that they are good enough for most purposes. The DHT22/AM2302 devices match the SHT71 well. SHT71 and E differ by a constant 0.4°C, the biggest divergence I have seen. Most of the others differ by ∼0.1°C. This is all consistent with my previous DHT22/AM2302 only results. The DHT11 does exibit more scatter but I have normally seen the specification given as ±2°C and my test device delivered ±0.7°C. I have elsewhere looked at absolute thermometer accuracy of the BME280, but that test has not been applied to these sensors.

Figure 9. Comparison of temperature outputs from the various device. I have no reason to suspect any problems with the thermometers in any of the devices.

Conclusion

If you have comments or suggestions feel free to contact me. robert -AT- kandrsmith.org


References: See here

Acknowledgements: See here

Almost vaguely related is a test I have started monitoring radon in the house too.